As Its Full Name Implies, Iec 62353 Medical Electrical Equipment – Recurrent Test and Test After Repair of Me Equipment, Defines the Requirements for...
As its full name implies, IEC 62353 Medical Electrical Equipment – recurrent test and test after repair of ME equipment, defines the requirements for electrical safety testing of medical electrical (ME) equipment and systems during routine intervals.
What is IEC medical device?
IEC 60601 is a series of technical standards that ensure the safety of medical electrical equipment. … It is assumed that public health authorities in many countries recognise IEC 60601-1 (Edition 3.2) as a pre-requisite for the commercialisation of electrical medical equipment.
What is the maximum ground resistance allowed in a medical device?
The grounding resistance is kept below approximately 0.15 ohms (set by local regulations) and is usually checked during preventative maintenance, prior to first use of the medical device and after each repair. A patient’s skin contact resistance can vary between 60,000 ohms to a few thousand ohms.